700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA — recall
Consumer goods · Australia · 2024-07-31
Hazard: The plunger nut threads on the calibrat…
700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA was recalled in Australia on 2024-07-31 due to the plunger nut threads on the calibration manifold may fai…. High-pressure pneumatic calibration manifold units included in 700HPPK test kits. All units with serial numbers below 6516999 are affected by this recall. Units with serial number 6517000 and above are not affected by this recall.
→ See all Fluke Australia recalls
| Country | Australia |
|---|---|
| Category | Consumer goods |
| Published | 2024-07-31 |
| Manufacturer | Fluke Australia |
| Distributor | RS Components Pty Ltd_x000D_ QLD Calibrations_x000D_ Triple Point Calibrations Pty Ltd_x000D_ Rushmore Distributors Pty Ltd_x000D_ Lawrence & Hanson_x000D_ Mektronics Australia Pty Ltd_x000D_ Bennic Calibration And Testing Services_x000D_ Thermo Fisher Scientific Australia_x000D_ Leda Electronics Pty Ltd_x000D_ Colterlec Pty Ltd |
| Model identifiers | Affected models: 700HPPK-NPT – Fluke No. 4784838 700HPPK-BSP – Fluke No. 4810462 700HPPK-MET – 4810443 |
| Hazard description | The plunger nut threads on the calibration manifold may fail, causing the plunger assembly to be ejected from the manifold at high speed. There is a risk of serious injury to the user if they are struck by the ejected plunger assembly. |
| Remedy | Follow this link to register your recalled manifold on the Fluke website. Please include your full contact information, including an email and shipping address._x000D_ Once registered, you will receive a return postage label so that it can be returned to FLUKE for repair and returned to you free of charge._x000D_ For more information, contact Fluke on 1300 135 853 (Monday to Friday, 8:30 am – 5:00 pm AEST) or by emailing ordersanz@fluke.com |
| Recall number | 2024/20252 |
| Official notice | View source ↗ |
Stay ahead of recalls like this
Monitor recalls across the US, EU, UK, Japan, Korea and more — with instant email alerts when a product like yours (or from your supplier) is recalled. Free to start.
Search & set alerts — free →Related consumer goods recalls
- The 2026 $2 Uncirculated Coin – Tooth Fairy KitAustralia · 2026-07-07
- Apeks XL4, XL4+ and Ocea scuba diving second-stage regulatorAustralia · 2026-07-06
- 8.5" LCD writing tabletAustralia · 2026-07-06
- 2 Ton Low Profile Fiernero Branded Hydraulic Trolley JackAustralia · 2026-07-03
- 12" LCD writing tabletAustralia · 2026-07-03
- Courant Kalimba arborist rope rangeAustralia · 2026-07-03