Recall365
RecallsEuropean Union › TRANS MEDIKAL - GETINGE Steam generator

TRANS MEDIKAL - GETINGE Steam generator — recall

Consumer goods · European Union · 2025-05-09
Hazard: Burns

TRANS MEDIKAL - GETINGE Steam generator was recalled in European Union on 2025-05-09 due to burns. Steam generator integrated into steriliser.

CountryEuropean Union
CategoryConsumer goods
Published2025-05-09
Model identifiers20 kW ; 22,5 kW ; 30 kW ; 37,5 kW ; 40 kW ; 45 kW ; 50 kW ; 60 kW
CategoryPressure equipment/vessels
HazardBurns
Hazard descriptionDue to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns.
Non-compliancePressure Equipment Directive
RemedyRecall of the product from end users. Making the marketing of the product subject to prior conditions.
Recall numberOR/00067/25
Official noticeView source ↗