700HPM High Pressure Calibration Manifold — recall

Consumer goods · United Kingdom · 2024-07-31
Hazard: Injuries

700HPM High Pressure Calibration Manifold was recalled in United Kingdom on 2024-07-31 due to injuries. 700HPM High Pressure Calibration Manifold.

CountryUnited Kingdom
CategoryConsumer goods
Published2024-07-31
CategoryPressure equipment / vessels
HazardInjuries
Hazard descriptionThis product presents a serious risk of injuries as during manual pressurization of the equipment, the plunger nut threads of the High-Pressure Pneumatic Calibration Manifold may yield releasing the plunger assembly and allowing it to be ejected at high velocity. The threads securing the plunger assembly have been found to fail prematurely. This can result in serious injury to the user.
Non-compliancePressure Equipment (Safety) Regulations 2016
RemedyFluke UK Ltd has stopped shipping the affected manifold and recalled the product from end users. Further information is available on the Fluke website at the following link: https://www.fluke.com/en-gb/support/safetynotices/fluke-700hxx-recall.
Recall number2407-0092
Official noticeView source ↗